1.1 This test method covers a procedure for the fatigue testing of metallic femoral hip
prostheses used in hip joint replacements. This test method covers the procedures
for the performance of fatigue tests on metallic femoral hip stems using a cyclic,
constant-amplitude force. It applies to hip prostheses that utilize proximal metaphyseal
fixation and are of a modular construct, and it is intended to evaluate the fatigue
performance of the modular connections in the metaphyseal filling (that is, proximal
body) region of the stem.