-
a) Type A: low-voltage and high-voltage, non-attenuating probe assemblies. Non-attenuating
probe assemblies that are rated for direct connection to voltages exceeding 30 V r.m.s., 42,4 V peak, or 60 V d.c.,
but not exceeding 63 kV. They do not incorporate components which are intended to
provide a voltage divider function or a signal conditioning function, but they may
contain non-attenuating components such as fuses (see Figure 1.)
-
b) Type B: high-voltage attenuating or divider probe assemblies. Attenuating or divider
probe assemblies that are rated for direct connection to secondary voltages exceeding 1 kV r.m.s. or 1,5 kV d.c.
but not exceeding 63 kV r.m.s. or d.c. The divider function may be carried out wholly
within the probe assembly, or partly within the test or measurement equipment to be
used with the probe assembly (see Figure 2).
-
c) Type C: low-voltage attenuating or divider probe assemblies. Attenuating or divider
probe assemblies for direct connection to voltages not exceeding 1 kV r.m.s. or 1,5 kV
d.c. The signal conditioning function may be carried out wholly within the probe assembly,
or partly within the test or measurement equipment intended to be used with the probe
assembly (see Figure 3).
-
d) Type D: low-voltage attenuating, non-attenuating or other signal conditioning probe
assemblies, that are rated for direct connection only to voltages not exceeding 30 V r.m.s., or 42,4 V peak,
or 60 V d.c., and are suitable for currents exceeding 8 A (see Figure 4).