This clause of Part 1 is applicable except as follows:
Replacement:
This Part 2 is applicable to electrically powered laboratory centrifuges.
NOTE If all or part of the equipment falls within the scope of one or more other Part 2 standards of IEC 61010 as well as within the scope of this standard, it will also need to meet the requirements of those other Part 2 standards.
Addition:
Add the following new item:
IEC 60034 (Rotating electrical machinery);
Addition:
Add the following five new items:
contact with moving parts (see 7.2);
laboratory centrifuge movement during any disruption (see 7.3.101);
high energy chemical reaction after rotordisruption (see 7.6.101.2 L));
ineffectiveness of bioseals (see 13.101)
Addition:
Add the following two new items:
additional precautions which may need to be observed when centrifuging materials which are flammable or explosive (see 5.4.101);
additional precautions which may need to be observed when centrifuging materials that could react chemically with sufficient vigour to cause a hazard (see 5.4.101).
Modification:
Modify item c) as follows:
temperature 2 °C to 40 °C;
Modification:
Modify item c) as follows:
ambient temperatures below 2 °C or above 40 °C;
EN 61010-2-020:2006
IEC 61010-2-020:2006