1.1.1 Equipment included in scope
Replace the existing text with the following:
This part of IEC 61010 specifies safety requirements for
hand-held and hand-manipulated current sensors described
below.
These current sensors are for measuring, detecting or injecting current, or
indicating current waveforms on circuits without physically opening the current
path of the circuit being measured. They can be stand-alone current sensors or
accessories to other equipment or parts of combined equipment (see Figure 101). These include measurement
circuits which are part of electrical test and measurement equipment, laboratory
equipment, or process control equipment. These current sensors and circuits need
additional protective means between the current sensor, the circuit and an
operator.
NOTE 1 Combined equipment is equipment that is electrically connected to a current
sensor by means of a permanent connection which can be detached only by the
use of a tool.
NOTE 2 Some current sensors are also known as current clamps, clamp
multimeters and current probes.
Current sensors are hand-manipulated before and/or after a test or measurement,
but do not necessarily need to be hand-held during the test or
measurement. Current sensors used as fixed equipment are not within the
scope of this document.
The following types of current sensors are covered:
-
a) Type A: a current sensor designed to be applied to or removed from
hazardous live uninsulated
conductors. Type A current sensors have
defined hand-held or hand-manipulated parts providing
protection against electric shock from the conductor being measured, and
also have protection against short-circuits between wires and between
busbars during clamping.
-
b) Type B: a current sensor which has protection against short-circuits
between wires or busbars during clamping but without defined
hand-held or hand-manipulated parts which provide
protection against electric shock during clamping. Additional protective
means are necessary to avoid electric shock from hazardous live
conductors which cannot be de-energised during application or removal of
the current sensor.
EXAMPLE 1
Flexible current sensors.
-
c) Type C: a current sensor without protection against short-circuits
between wires or busbars during clamping. Type C current sensors are
intended to be applied to or removed from hazardous live
uninsulated conductors or from
non-limited-energy circuit conductors only when they are
de-energised.
EXAMPLE 2
Split-core transducers.
-
d) Type D: a current sensor designed to be applied to or removed from
insulated conductors or from limited-energy circuit conductors.
EXAMPLE 3
Current probes for oscilloscopes and earth leakage current
detectors.
All current sensors can also be used with insulated conductors. In this case,
hazards are limited to acceptable levels by the insulation of the
conductors.
Additional requirements for clamp multimeters are
given in Annex EE.
Figure 101 shows graphical
representations of typical current sensors for illustration purposes. Current
sensors can look different depending on the design.
Figure 101
—
Examples of current sensors and their parts
|
|
|
|
Type A current sensor as an accessory
|
Type A current sensor with self-contained and/or
additional measuring functions
|
|
|
|
|
Type A current sensor with fork-style
jaws
|
Type B flexible current sensor part of a piece of
combined equipment
|
|
|
|
|
Type C split-core current sensor
|
Type D current sensor for non-hazardous live
applications (shown with a sliding jaw)
|
| 1 |
Jaw end(S) |
| 2 |
Jaw
|
| 3 |
Measuring circuit
terminals
|