1 Scope
This document applies to focal spots in medical diagnostic X-ray tube assemblies for medical use, operating at X-ray tube voltages up to and including 150 kV.
This document describes the test methods employing digital detectors for determining:
-
a)
focal spot
dimensions in terms of nominal focal spot values, ranging from 0,1 to 3,0;
-
b)
line spread functions;
-
c) one-dimensional modulation transfer functions;
-
d)
focal spot pinhole radiograms,
and the means for indicating compliance.
In informative annexes, star pattern imaging and blooming value are described.